Inside the TOMY Autoclave ES-215/315 is subject to very high temperature and pressure conditions during operation.

Manufacturer:
TOMY
Availability:
In Stock
SKU:
ES-215-ES-315

Inside the TOMY Autoclave ES-215/315 is subject to very high temperature and pressure conditions during operation. Therefore, should the autoclave be incorrectly installed or operated, there might be potentially hazardous situation that could cause death or serious injury to the operator and laboratory personnel or damage to property. Read this operator’s manual thoroughly for a complete understanding of the autoclave and its correct handling before operating the instrument.

 

Specifications

Model ES-215 ES-315
Operating temperature range
(during sterilizing)
105 - 123°C 105 - 132°C
Operating pressure range 0 - 127kPa 0 - 186kPa
Maximum operating pressure 147kPa 216kPa
Temperature control Digital, microprocessor controlled
Temperature display / display range Digital / -15 - 180°C
Pressure display / display range Analog / 0 - 250kPa Analog / 0 - 400kPa
Heat source 1.5kW electric heater 2.0kW electric heater
Timer control Digital, microprocessor controlled
Timer display Digital
Timer
display range
Sterilizing mode: 1 - 240min. or continuous
Sterilizing-warming mode: 1 - 240min. (sterilizing) / 1 - 8hours (warming)
Heating mode: 1 - 8hours
Start-timer setting range 1 - 99hours in each operation mode
Memory function A set of temperature and time setting in each operation mode
Safety devices • Inside the chamber overheat prevention, • Outer wall of the chamber overheat prevention, • Overpressure prevention, • Temperature sensor disconnection prevention, • Empty heating prevention, • Leakage breaker, • Safety valve
Malfunction prevention devices • Low water level detection, • Exhaust valve knob open/close detection,
• Insufficient sterilization detection, • Chamber lid open/close detection
Leakage
breaker
Rated breaking current: 20A (120V) , 15A (220 / 230 / 240V) 30A (120V) , 15A (220 / 230 / 240V)
Rated sensed current: 30mA (120V) , 10mA (220 / 230 / 240V)
Protection type Class 1 equipment
Operating
environment
Ambient temperature: 5 - 35°C
Relative humidity: 30 - 85%
Atmospheric pressure: 700 - 1060hPa
Gradient: Within 2°
Dimensions φ248 x 543mm φ325 x 740mm
Usable dimension φ236 x 450mm φ312 x 635mm
Capacity of chamber 22ℓ 53ℓ
Chamber material SUS304
Capacity / material of exhasut bottle 3ℓ / polyethylene
Type / material of lid gasket Accommodating internal pressure / silicone rubber
Dimensions of main unit (mm) 400W x 460D x 920H (Height from floor to control panel:705) 490W x 560D x 1090H (Height from floor to control panel:875)
Net weight 50kg 80kg
Rated Voltage AC 120 / 220 / 230 / 240V
Power input 15A (120V), 7A (220 / 230 / 240V) 20A (120V), 10A (220V), 9A (230 / 240V)
Power supply requirements Single-phase 120V AC (50 / 60Hz) 15A or above Single-phase 120V AC (50 / 60Hz) 20A or above
Single-phase 220 / 230 / 240V AC (50 / 60Hz) 15A or above
Power Consumption (calorific value) 1.5kW (1290kcal/h) 2kW (1720kcal/h)
Accessories • Stainless steel basket with slatted bottom plate 1, • Water level sensor 1,
• Inspection sheet 1, • Warranty card 1, • Customer card 1, • Operator's manual 1,
• Clear plastic accessory case 1, • Attachment screw for clear plastic accessory case 1
Stainless basket 1 (φ224 x 376mm) Stainless basket 2 (φ300 x 182mm)